Block of detection BDKN-200 is intended  for  measuring the power of the ambient dose equivalent Ḣ*(10) of neutron radiation and for measuring the neutron flux density.

Output signal  corresponds to the standard interface:
• RS – 485 – detection unit BDKN-200 PD;

• RS – 422 – detection unit BDKN-200 DD;
The contents and the format of the data are defined in the report of the informational exchange DiBUS (www.doza.ru)

1.Mode of operation: continuous or switching on and off without limiting the number of switching on and duration of use in ON state.
2.Energy range of detected neutron radiation is from 0,025 eV to 10.0 MeV.
3.The measuring range of PED of Ḣ * (10) of neutron radiation is from 0,1 mSv / h to 0.1 Sv / h.
4.Range of measurement of neutron flux density is from 0,1 up to 104 neutron / (c • cm2).
5.Limit of the basic relative error of the block is  ± 30%.
6.Sensitivity of the detection unit is 5,0 imp/s at 1,0 mkЗv/h
7.Energy dependence of the block in the neutron energy range of typical neutron spectra is no more than 30%.
8.Time of setting up the operating mode is no more than 5 min
9.Duration of continuous work of the unit is not less than 24 hours
10.Algorithm for processing measuring data is a method of “moving average”. Integration time is 40 seconds. Exposure – 1.
11.Dimensions of the unit: diameter is 40 mm, length is 250 mm.
12.Weight of the block is no more than 0,5 kg.

ELECTRICAL CHARACTERISTICS

1. Power supply – is DC voltage from 7 to 42,0 V.

2. Consumed current is not more than 10 mA.

Designation of the connector’s contacts (plug 2PMGD24B10SH5) , block BDKN-200PD (RS-485).

STRUCTURE OF DATA REGISTERS

To see the structure of data DiBAS-register, please refer to WWW.TETRA.UA

Designation of the connector’s contacts (plug 2PMGD24B10SH5) , block BDKN-200PD (RS-485).

Contact

Signal designation

Commentary

1

+ Uп

Positive supply voltage 7 – 24 В

4

DTR+

Data transceiver’s output is direct

5

DTR-

Data transceiver’s output is inverse

9

Casing/frame

Ground connection

10

General

Negative supply voltage

Designation of the connector’s contacts (plug 2PMGD24B10SH5) , block BDKN-200DD (RS-422).

Contact

Signal designation

Commentary

1

+ Uп

Positive supply voltage 7 – 24 В

4

Tx+

Data transmitter’s output is direct

5

Tx-

Data transmitter’s output is inverse

6

Rx+

Data receiver’s input is direct

8

Rx-

Data receiver’s input is inverse

9

Casing/frame

Ground connection

10

General

Negative supply voltage
Další Produkty